S.M. Scientific Instruments (P) LTD

Class II Type A2 Biological Safety Cabinet SMI-163A-4

Home Clean Air >Biosafety Cabinets > Class II Type A2 Biological Safety Cabinet SMI-163A-4

Class II Type A2 Biological Safety Cabinet SMI-163A-4

CLICK TO SCAN

Description

JINDAL Biological Safety Cabinet Class II type A2 is vertical designed with freestanding configuration model stand with castors to protect the lab staff, the environment and sensitive work processes in which biological agents are used. The safety cabinet offers a high level of ULPA/HEPA filter with 99.999% efficiency at 0.1µm to 0.3 µm particle & an airflow pattern of 70% air recirculation and 30% exhaust. Stand height is 28 inches.

The Cabinet construction made of galvanized steel with powder coating material. The interior of the cabinet including the work area and side panels and back panel are made of stainless steel 304 material.

The Cabinet is equipped with a smart, safe and user-friendly touch-screen controller display system that protects the operator and provides alerts for periodic maintenance actions and parts’ replacement. The cabinet has safety reasons for the user & the environment and accessories is supplied with each cabinet Foot operated burner, canopy to connect external duct.

The cabinet is precisely control ventilation to sweep airborne particles produced during routine microbiological procedures way from the user while maintaining a controlled environment for the specimen. The Bio safety cabinet has consistent and reliable down flow velocity across the supply HEPA filter over the life of the cabinet. The cabinet has a pressure sensor for detect pressure drop across the supply filter, rather than at just one point across the down flow. Pressure sensor encased to protect the sensor from temperature, humidity and other environmental phenomena that impact the sensor’s performance. The cabinet has stable and self-compensating airflow, despite building voltage fluctuations & filter loading.

The system also has a programmable “standby” mode that shuts down all unnecessary electricity consumption and sets vital components at the required safety level. The safety cabinet has dual DC direct current electronically commutated motor automatically adjust the airflow speed without the use of a damper to ensure continuous safe working conditions.

Transparent toughened sash with conventional weight counter balance system.

Detailed Images

Advantages

  • EN12469 certified
  • H14 as per EN-1822 certified Filter
  • Dual stage filtration system
  • Low heat & low energy consumption
  • Standby mode to economize power consumption
  • Economy mode of operation to reduce energy consumption
  • Vertical airflow direction
  • Microprocessor control of air velocities in real-time with the display
  • Class 2 air cleanliness class as per ISO 14644-1
  • Cabinet have levelling jacks
  • UV lamp for decontamination
  • Advanced air pressure sensing technology
  • Display for monitoring the condition of airflow / air velocities in real time, UV light on / off , white light on / off , UV and HEPA filter usage hours , remaining life of the filters / alarm , filter change
  • Smooth interior and exterior
  • Blower and filtration facility
  • Pressure monitoring system with pressure drop 13-14 mm
  • Airflow sensor to alert the user if airflow sufficient for greater safety
  • UV Lamp with settable timer and interlock for safety
  • Indicator or buzzer mechanism, activating when the front window is opened beyond the recommended level
  • UV light programmable to allow for specific exposure times from 0 to 24 hours
  • UV light fitted at appropriate position for disinfecting the work area
  • Removable work plates
  • Auto speed increase on the opening door for maintaining velocity
  • Automatic component features that adjust blower speed to maintain the optimum airflow in case of filter loading
  • Front glass accessible for cleaning without requiring the user remove or support the glass window
  • RS –232 USB port system available
  • 2 Nos. of LED light with brightness level adjustment
  • Work on 220-240V, Single Phase 50Hz power supply

Additional Accessories

  • HEPA filter
  • Stainless steel arm rest
  • Magnehelic gauge
  • Gas burner
  • Electrical outlet (2 Nos.)

Technical Data

Specs/Model

SMI-163A-3

SMI-163A-4

SMI-163A-6

External Size (WxDxH)

1030 x 775 x 2295mm

1380 x 775 x 2295mm

1870 x 775 x 2295mm

Internal Size (WxDxH)

900 x 600 x 600 mm

1200 x 600 x 600 mm

1700 x 600 x 600mm

Working Area Height

750mm

750mm

750mm

Maximum Front Sash Opening

520mm

520mm

520mm

Tested Front Sash Opening

200mm

200mm

200mm

Production/ Test Standard

CE / In accordance with EN12469

CE / In accordance with EN12469

CE / In accordance with EN12469

Air Flow Volume (Inflow)

417 CFM

417 CFM

417 CFM

Air Flow Volume (Downflow)

663 CFM

663 CFM

663 CFM

Air Flow Volume (Exhaust)

459 CFM

459 CFM

459 CFM

Downflow Velocity

0.30m/s, 60FPM

0.30m/s, 60FPM

0.30m/s, 60FPM

Inflow Velocity

0.53m/s, 105FPM

0.53m/s, 105FPM

0.53m/s, 105FPM

Front Window 10” sash open

5mm toughened safety glass sash

5mm toughened safety glass sash

5mm toughened safety glass sash

Filter

ULPA/HEPA filter with 99.999% efficiency at 0.1µm to 0.3 µm particle

ULPA/HEPA filter with 99.999% efficiency at 0.1µm to 0.3 µm particle

ULPA/HEPA filter with 99.999% efficiency at 0.1µm to 0.3 µm particle

Noise Level

63 dB

63 dB

63 dB

Internal Glare- free Fluorescent light

1200Lux Eco-friendly

1200Lux Eco-friendly

1200Lux Eco-friendly

Display

LED Display with Pressure Gauge Differential

LED Display with Pressure Gauge Differential

LED Display with Pressure Gauge Differential

Power Sockets

Two (15/5 Amp)

Two (15/5 Amp)

Two (15/5 Amp)

UV Light

30W ; 254nm

30W ; 254nm

30W ; 254nm

Audio-Visual Alarms

Deviation in airflow velocity,

Filter replacement, sash

position, UV life timer, blower failure, window position

Deviation in airflow velocity,

Filter replacement, sash

position, UV life timer, blower failure, window position

Deviation in airflow velocity,

Filter replacement, sash

position, UV life timer, blower failure, window position

Taps

Water Tap : 1

Gas Tap : 1

Vacuum/Air Tap : 1

Water Tap : 1

Gas Tap : 1

Vacuum/Air Tap : 1

Water Tap : 1

Gas Tap : 1

Vacuum/Air Tap : 1

Weight

316Kg

371Kg

390Kg

 

Product Brochure

Product Brochure Link here