Home ≫Clean Air >Biosafety Cabinets >Class II Type A2 Biological Safety Cabinet SMI-163MS
JINDAL Biological Safety Cabinet Class II type A2 is vertical designed with freestanding model stand with castors to protect the lab staff, the environment and sensitive work processes in which biological agents are used. The safety cabinet offers a high level of ULPA filter with filtration efficiency 99.999% for 0.1-micron particles & an airflow pattern of 70% air recirculation and 30% exhaust. Stand height is 28 inches.
The Cabinet consist work bench construction with Stainless steel 304 grade material. The outer body made of epoxy coated steel material removable, seamless dished work surface with lift out knobs. The interior of the cabinet including the work area and side panels and back panel are made of stainless steel 304 material.
The Cabinet is equipped with a smart, safe and user-friendly HMI touch-screen control system that protects the operator and provides alerts for periodic maintenance actions and parts’ replacement. The cabinet has safety reasons for the user & the environment and accessories is supplied with each cabinet Foot operated burner, canopy to connect external duct.
The cabinet is precisely control ventilation to sweep airborne particles produced during routine microbiological procedures way from the user while maintaining a controlled environment for the specimen. The Bio safety cabinet has consistent and reliable down flow velocity across the supply HEPA filter over the life of the cabinet. The cabinet has a pressure sensor for detect pressure drop across the supply filter, rather than at just one point across the down flow. Pressure sensor encased to protect the sensor from temperature, humidity and other environmental phenomena that impact the sensor’s performance. The cabinet has stable and self-compensating airflow, despite building voltage fluctuations & filter loading.
The system also has a programmable “standby” mode that shuts down all unnecessary electricity consumption and sets vital components at the required safety level. The safety cabinet has dual DC motor automatically adjust the airflow speed without the use of a damper to ensure continuous safe working conditions.
Transparent toughened sash with Conventional weight counter balance system.
Specs/Model | SMI-163MS (I) | SMI-163MS (II) | SMI-163MS (III) |
External Size (W x D x H) | 1030 x 750 x 2130mm | 1360 x 750 x 2130mm | 1870 x 750 x 2130mm |
Internal Size (W x D x H) | 900 x 600 x 600 mm | 1200 x 600 x 600 mm | 1700 x 600 x 600mm |
Working Area Height | 500mm | 500mm | 500mm |
Maximum Front Sash Opening | 254mm | 254mm | 254mm |
Tested Front Sash Opening | 175mm | 175mm | 175mm |
Production/ Test Standard | CE / In accordance with EN12469 | CE / In accordance with EN12469 | CE / In accordance with EN12469 |
Air Flow Volume (Inflow) | 350 CFM | 350 CFM | 350 CFM |
Air Flow Volume (Downflow) | 225 CFM | 225 CFM | 225 CFM |
Air Flow Volume (Exhaust) | 350 CFM | 350 CFM | 350 CFM |
Downflow Velocity | 0.33m/s, 60FPM | 0.33m/s, 60FPM | 0.33m/s, 60FPM |
Inflow Velocity | 0.54m/s, 105FPM | 0.54m/s, 105FPM | 0.54m/s, 105FPM |
Front Window 10” sash open | 5mm toughened safety glass sash | 5mm toughened safety glass sash | 5mm toughened safety glass sash |
Filter | ULPA filter with filtration efficiency 99.999% for 0.1-micron particles | ULPA filter with filtration efficiency 99.999% for 0.1-micron particles | ULPA filter with filtration efficiency 99.999% for 0.1-micron particles |
Noise Level | 58 dB | 58 dB | 58 dB |
Internal Glare- free Fluorescent light | 1200Lux Eco-friendly | 1200Lux Eco-friendly | 1200Lux Eco-friendly |
Display | LED Display with Pressure Gauge Differential | LED Display with Pressure Gauge Differential | LED Display with Pressure Gauge Differential |
Power Sockets | Two (15/5 Amp) | Two (15/5 Amp) | Two (15/5 Amp) |
UV Light | 30W ; 254nm | 30W ; 254nm | 30W ; 254nm |
Audio-Visual Alarms | Deviation in airflow velocity, Filter replacement, sash position, UV life timer, blower failure, window position | Deviation in airflow velocity, Filter replacement, sash position, UV life timer, blower failure, window position | Deviation in airflow velocity, Filter replacement, sash position, UV life timer, blower failure, window position |
Taps | Water Tap : 1 Gas Tap : 1 Vacuum/Air Tap : 1 | Water Tap : 1 Gas Tap : 1 Vacuum/Air Tap : 1 | Water Tap : 1 Gas Tap : 1 Vacuum/Air Tap : 1 |
Weight | 316Kg | 371Kg | 390Kg |